Bowie-Dick Test Pack, Sterile Monitoring

Still deciding? Get samples of $ !
Request Sample

Product Description

Basic Information

Model NO.
TOPBD115125
Sterilization
Without Ethylene Oxide
Guarantee Period
Three Years
Group
Hospital
Logo Printing
With Logo Printing
Transport Package
100PCS/CTN
Specification
115*125mm
Trademark
OEM
Origin
Anqing
HS Code
3005101000
Production Capacity
500,000 PCS/Year

Product Description

B-D Test Pack

1. Material: Paper with Steam Indicator

Bowie-Dick Test Pack Material

2. Indicator Performance:

Steam Sterilization: 121°C (15min) / 134°C (3.5min)
Result: Yellow changes to Black
Bowie-Dick Test Pack Indicator

3. Key Features:

  • More accurate, compact, environmentally friendly and cost effective than traditional Bowie Dick tests.
  • Test for both air leaks and wet steam content.
  • Detect air leaks in a pre-vacuum steam sterilizer at 270°F (132°C) at a minimum 3.5-minute exposure.
  • Distinctive color change is easy to read and not reversible.
  • Lead-free ink and no toxic heavy metals.

4. Size & Specifications:

Code Size Pack
TOPBD115125 115cm*125cm 100pcs/ctn

5. Customization: We provide production for customer brands, specific sizes, and custom designs.

6. Delivery Time: Within 15 days.

Frequently Asked Questions

What is the primary purpose of the B-D Test Pack?
The B-D (Bowie-Dick) Test Pack is designed to detect air leaks and evaluate the air removal and steam penetration capabilities of pre-vacuum steam sterilizers.
How do I interpret the test results?
A successful test is indicated by a uniform color change of the indicator ink from yellow to black. Any non-uniform color change suggests a potential issue with the sterilizer.
Is the indicator ink safe for the environment?
Yes, our B-D Test Packs use lead-free ink and contain no toxic heavy metals, making them environmentally friendly and safe to handle.
Can the color change be reversed after the test?
No, the distinctive color change is irreversible, providing a permanent record of the sterilization monitoring process.
What is the shelf life of these test packs?
The quality guarantee period for our B-D Test Packs is three years when stored under proper conditions.
Do you offer OEM or custom branding services?
Yes, we can produce according to customer requirements, including custom branding, specific sizes, and unique designs.